The Multiwave photometer PIR3502 is designed for the IR (infrared) and NIR (near infrared) spectral regions.  The ABB Multiwave photometers provide continuous measurements in complex streams which allow for fast reactions to process changes. The fast analysis provides optimum control of a process and improves its safe operation. Measurements can be made in liquid or vapor streams.




.

Multiwave Photometers

Overview
Applications
Specifications
Documentation
Service

Overview

Infrared Photometer PIR3502

The PIR3502 Multiwave process photometer provides continuous measurements of vapor or liquid samples, in simple or complex process streams, for process control, product quality assurance, safety, catalyst protection and area monitoring.

The filter wheel of the photometer utilizes up to eight wavelengths, increasing the number of measurement solutions. Because it uses multiple wavelengths, the photometer can compensate for many types of interference and handle multiple component applications.

The PIR3502 Multiwave photometer provides the following:

Continuous measurements
Multiple component samples
Measures Vapor or Liquid samples
Operates in IR and NIR
Fiber optic option for NIR applications
Multiple interference compensation capability
VistaNET and VistaSTAR Connectivity.

 

Applications

Infrared Photometer PIR3502

Multiwave Photometer Solutions

The ABB PIR3502 process photometers leading applications include:

Infrared

Isocyanate in chloroaromatic solvent
Ambient air monitoring
Multicomponent monomers in polymer process
With Infrared (IR) Spectral Regions: 2.50 - 14.5 um

Near Infrared

Measurement of Hydrogenic Compounds (C-H, N-H, O-H)
Water in Organics, such as Ethylene Dichloride (EDC)
Caustic in Acid Gas Scrubbers
With Near Infrared (NIR) Spectral Regions: 800 - 2500 nm

 
 
 
 

Specifications

Performance Specifications

Precision: ± 1 % of full scale
Noise: ± 1 % of full scale at 0.02 absorption units
± 0.5 % of full scale at 0.20 absorption units
Linearity: Standard ± 2 % of full scale
Zero Drift: ± 0.5 % of full scale per day for IR;
   

Operating Specifications

Ambient Temperature Range: 32° to 113°F (0--45°C)
Sample Pressure: 0--500 PSIG (0--34 bar)
Voltage Input Variation: 10 % fluctuation without causing an output variation of 0.05 % of full scale
Electrical Power to Analyzer: 100/115/220/230 VAC, 45--66 HZ; 150W, Maximum Power Consumption; 600W With Electric Heat
   

General Installation

Protect the instrument from direct sunlight and rain at operating temperature between 32°--113° F (0--45°C)
   

Dimensions

Weight: approximately 80 lbs (36.28 kg)
Height: 13.5 inches (342.9 mm)
Depth: 10.5 inches (266.7 mm)
Width: 10 inches (254 mm) per module
Overall Length: (based on cell pathlength)
Minimum: 26.8 inches (681.4 mm) for 0.5 to 16 mm cell
Maximum: 65.5 inches (1665 mm) for 1 meter cell
   

Area Classification

Standard: NEC Class 1, Division 2, Group B, C, D
Enclosure Purge: NEC Class 1, Group B, C, D, Division 1, with "Y"
Purge CENELEC Zone 1, IIB + H2